Animal Fit V6 Infusion Pump
Configuration |
Applicable Infusion Sets |
1ml/15d, 1ml/19d, 1mW20d, 1ml/60d |
Flow Rate |
Ml/h mode: 0.1 . 1200ml/h
(in 0.01ml/h increments) .
drop/min mode: 1 .(400ml/h*d/60) d/min
(in 1d/min increments) |
BOLUS |
Bolus rate: 0.1 . 1200mlh (in 0.01mlsh increments) .
VTBI( Volume to be infused): 0.1 . 100ml (in 001mlincrements) |
FF(Fast Forward) |
FF rate: 01 1200ml/h (in 0.01ml/h increments) .
VTBI( Volume to be infused): 0 - 9999.99ml (in 0.01ml increments) |
VTBI |
0.1 . 9999.99ml (in 0.01mlincrements) |
Cumulative Infusion Volume |
0. 9999 .99ml (in 0.01mlincrements) |
Rate (ml/h) Accuracy |
≤±5%(after calibration): |
Rate (d/min) Accuracy |
≤±5% |
Occlusion Pressure |
5 occlusion pressure levels (20,50,80,110,130)kPa |
Alarms Information |
Near infusion end, infusion finished, standby task finished, occlusion alarm.KVO completed, door opening, air in-line, forgotten operation,low battery, battery empty, abnormal running state, equipment failure.
power failure, replace tube, drop sensor is pulled out, Infusion set mismatch.impending occlusion(optional), empty bottle, battery disconnected etc. |
Dimensions |
147mm (L) x195mm (W) x235mm (H) |
Contact via WhatsApp